The datalog and treatment tip from the patient event were returned and evaluated.Errors had occurred during treatment in the log.Errors indicate a recoverable problem that requires operator intervention.If the error occurs during a radio-frequency (rf) treatment, the rf delivery will be stopped, then a post-cooling step will be completed prior to generating an ¿error tone¿ and an event code and message will be displayed.After the error tone, the system will transition into action required mode and will display text with instructions for the operator indicating what action may be required to resolve the issue.Based on the evaluation of data the system and handpiece performed as expected.The treatment tip passed flow and thermistor testing but had failed leak testing and visual inspection.Images of the tip confirmed damage to the tip membrane along the rf trace.Breakdown of the dielectric material can cause the rf energy, delivered by the system, to focus in a small area of the membrane, rather than to be uniformly distributed over the entire membrane area.Defects on the tip membrane can lead to a raise in temperature of the tip during treatment and can potentially cause patient burns.Damage to the rf trace can also cause the reported sparking noticed during treatment.Thermage cpt system technical user¿s manual instructs the operator to inspect the treatment tips for any signs of physical damage prior, during, and after treatment.With respect to all thermage systems clinicians should frequently inspect the tip membrane during treatment for signs of breakdown and build-up of foreign substances.Investigation found stress concentrations on the flex assembly at the adhesive edge that damaged the rf trace, causes arcing and subsequent burn-through of the flex circuit membrane.A review of the manufacturing records showed all requirements were met.Final test verification specifications are acceptable.No non-conformities or anomalies found related to this complaint when reviewing the device history record for the serial/lot number.Based on the available information, this event was most likely caused by damage on the tip membrane.
|