• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number MCP00706041#ROTAFLOW ENGLISH/UK UK-PLUG
Device Problem Increased Pump Speed (1501)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.Further information has been requested but has not yet been received.A follow up medwatch will be submitted when additional information becomes available.
 
Event Description
The event occurred in india.It was reported that the error message ¿run away error¿ occurred on the rotaflow console before use and no patient was involved.No harm to any person has been reported.Complaint id: (b)(4).
 
Manufacturer Narrative
The event occurred in india.It was reported that the error message ¿run away error¿ occurred on the rotaflow console before use and no patient was involved.No harm to any person has been reported.The rotaflow console with s/n (b)(6) was investigated by a getinge field service technician and the technician was able to reproduce the reported failure "run away error".The rf (rotaflow) control board pcba (printed circuit board assembly) kit (article number (b)(4)) and the rf flow measure pcba (article number (b)(4)) has been replaced.After the replacement the device is working as intended.The affected control and flow measurement board were requested for further investigation but both boards are not available for investigation.Based on the investigation results the reported failure "run away error" could be confirmed.The failure mode "run away error" can be linked to the following most possible root causes according to the rotaflow risk management file.Defective motor control electronics, pump stop intervention after technical error (e.G.Pump runaway, error head), software error, wrong intervention limits.The review of the non-conformities was performed on (b)(3) 2022 and during the period of (b)(3) 2019 to (b)(3) 2022 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The rotaflow console in question was produced in (b)(3) 2019.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.
 
Event Description
Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key14753233
MDR Text Key301915743
Report Number8010762-2022-00233
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMCP00706041#ROTAFLOW ENGLISH/UK UK-PLUG
Device Catalogue Number701043291
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-