CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE Back to Search Results
Model Number MCH-1000(I)
Device Problems Fluid/Blood Leak (1250); Smoking (1585); Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/30/2021
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is reporting the following complaint after a voluntary review of all complaints (reportable or not) since 2016.This report is being filed now, after being scrutinized under a newly revised risk matrix, recently adopted after inspection.The device was found to be contaminated with m.Abscessus (50 cfu/ml).However, the customer has ceased communication and an initial investigation was not possible due to the lack of access to the device.
 
Event Description
Customer sent water samples to an independent lab to find out what they would culture.After receiving the lab results back, it was determined that the device was contaminated with m.Abscessus (50 cfu/ml).
 
Event Description
Customer reports that their device started producing smoke and leaking during a case.
 
Manufacturer Narrative
Cardioquip's investigation determined that the device was operated without water in the tank.The damaged components were replaced; the device passed inspection and is fully operational.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOQUIP MODULAR COOLER HEATER
Type of Device
CARDIOPULMONARY BYPASS DEVICE
Manufacturer (Section D)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer (Section G)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer Contact
charley ford
8422 calibration ct.
college station, TX 77845
9796910202
MDR Report Key14755155
MDR Text Key294586451
Report Number3007899424-2022-00025
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102147
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Distributor
Reporter Occupation Other Health Care Professional
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMCH-1000(I)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age63 YR
Patient SexMale
Patient Weight75 KG
-
-