Model Number MCH-1000(I) |
Device Problems
Fluid/Blood Leak (1250); Smoking (1585); Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/30/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The manufacturer is reporting the following complaint after a voluntary review of all complaints (reportable or not) since 2016.This report is being filed now, after being scrutinized under a newly revised risk matrix, recently adopted after inspection.The device was found to be contaminated with m.Abscessus (50 cfu/ml).However, the customer has ceased communication and an initial investigation was not possible due to the lack of access to the device.
|
|
Event Description
|
Customer sent water samples to an independent lab to find out what they would culture.After receiving the lab results back, it was determined that the device was contaminated with m.Abscessus (50 cfu/ml).
|
|
Event Description
|
Customer reports that their device started producing smoke and leaking during a case.
|
|
Manufacturer Narrative
|
Cardioquip's investigation determined that the device was operated without water in the tank.The damaged components were replaced; the device passed inspection and is fully operational.
|
|