MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED2-475-20
Device Problems Unintended Movement (3026); Activation Failure (3270); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2022
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803. This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda. Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report. This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report. In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿. These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting. Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms. This statement should be included with any information or report disclosed to the public under the freedom of information act. Any required fields that are unpopulated are blank because the information is currently unknown or unavailable. A good faith effort will be made to obtain the applicable information relevant to the report. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received a report that the pipeline failed to open distally. The patient was undergoing treatment for an unruptured, fusiform aneurysm located in the superior hypotheseal. The max diameter was 7mm, and the neck diameter was 5mm. The patient's vessel tortuosity was severe. The landing zone was 4. 00mm distal and 4. 65mm proximal. It was reported that pipeline opened nicely in the hypotheseal segment but it fell down too proximal. The physician recaptured the device and moved it distally. The doctor began to push the device out of the microcatheter, and the device had a narrowing at the distal end that was evident under flouro. This caused malposition on the distal end of the device. The doctor removed the device, and a replacement was deployed successfully. The pipeline was not positioned in a bend, and more than 50% had been deployed when it failed to open. The pipeline had been resheathed less than 2 times. No additional steps were taken to open the pipeline.  the patient did not experience any injury or complications. Angiographic results post procedure showed successful deployment of the replacement device with no injury to the patient. The devices were prepared according to the instructions for use (ifu). Ancillary devices include a phenom 27 microcatheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePIPELINE FLEX W/SHIELD TECHNOLOGY
Type of DeviceINTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key14757013
MDR Text Key294421598
Report Number2029214-2022-01034
Device Sequence Number1
Product Code OUT
UDI-Device Identifier00763000284633
UDI-Public00763000284633
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 09/02/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberPED2-475-20
Device Catalogue NumberPED2-475-20
Device Lot NumberB374613
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-