MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD SAF-T-INTIMA¿ WITH Y ADAPTER YEL 24GA X 19MM; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD SAF-T-INTIMA¿ WITH Y ADAPTER YEL 24GA X 19MM; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383319
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/23/2022
Event Type  malfunction  
Event Description
It was reported while using bd saf-t-intima¿ with y adapter yel 24ga x 19mm the tubing separated from the connector.There was no report of patient impact.The following information was provided by the initial reporter: during use.We have had a problem with a 24g intimate, when clamping the tube of the connection has been disconnected.
 
Manufacturer Narrative
Our quality engineer inspected the 1 photo submitted for evaluation.The reported condition of separation adapter from tubing was not confirmed upon inspection of the photo.The reported issue of lie distance incorrect was confirmed upon inspection of the photo.However, bd cannot confirm the cause of the failures to our manufacturing process since no sample was returned for evaluation.Examination of the actual product involved may provide clarification as to the cause for the reported failure.A device history record review showed no non-conformances associated with these issues during the production of this batch.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAF-T-INTIMA¿ WITH Y ADAPTER YEL 24GA X 19MM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14777709
MDR Text Key301080096
Report Number9610847-2022-00230
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number383319
Device Lot Number1293549
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/17/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-