Device Problem
Material Puncture/Hole (1504)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/22/2022 |
Event Type
malfunction
|
Event Description
|
It was reported an unknown issue, possible damage to trach tube.After removing the trach from the water it was boiled in, it was discovered there was a tear on the side of it.Lot and part number are unknown.Bivona 4.0 flextend cuffed l41.Discovered prior to use, no patient involvement was reported.
|
|
Manufacturer Narrative
|
Device evaluation: the device was returned for investigation.Visual inspection and functional test were performed.Visual inspection found no defects with the device.Functional testing successfully passed, and the failure mode reported was not confirmed as the reported problem was not duplicated.No root cause was determined as the sample successfully passed functional testing.No corrective actions were taken as the complaint was not confirmed.This remediation mdr was generated under protocol b10010579, as a result of warning letter cms# (b)(4).A dhr (device history review) was performed, and no problems or issues were identified during this dhr review.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 if additional reportable information becomes available., corrected data: g2: manufacturing site address is unknown.D4: catalog number is unknown.D4: udi information is unknown.G5: premarket (510k) number is unknown.
|
|