MAUDE Adverse Event Report: NULL BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NULL BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Device Problem Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/22/2022
Event Type  malfunction  
Event Description
It was reported an unknown issue, possible damage to trach tube.After removing the trach from the water it was boiled in, it was discovered there was a tear on the side of it.Lot and part number are unknown.Bivona 4.0 flextend cuffed l41.Discovered prior to use, no patient involvement was reported.
 
Manufacturer Narrative
Device evaluation: the device was returned for investigation.Visual inspection and functional test were performed.Visual inspection found no defects with the device.Functional testing successfully passed, and the failure mode reported was not confirmed as the reported problem was not duplicated.No root cause was determined as the sample successfully passed functional testing.No corrective actions were taken as the complaint was not confirmed.This remediation mdr was generated under protocol b10010579, as a result of warning letter cms# (b)(4).A dhr (device history review) was performed, and no problems or issues were identified during this dhr review.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 if additional reportable information becomes available., corrected data: g2: manufacturing site address is unknown.D4: catalog number is unknown.D4: udi information is unknown.G5: premarket (510k) number is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section G)
NULL
MDR Report Key14778015
MDR Text Key294533874
Report Number3012307300-2022-12387
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/28/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-