MEDTRONIC NEUROMODULATION IMPLANTABLE NEUROSTIMULATOR; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number NEU_INS_STIMULATOR |
Device Problems
Disconnection (1171); High impedance (1291); Low impedance (2285); Impedance Problem (2950); Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: neu_unknown_lead, serial# unknown, product type: lead.Other relevant device(s) are: product id: neu_unknown_lead, serial/lot #: unknown.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient has bilateral stn dbs for parkinson's disease.The patient continues to do well with good symptom control.In 2017, noted high impedance readings on the right for all contacts so the patient was programmed in constant current mode and all has been well.The patient still has good symptom control, but recently the patient programmer gave an alert "out of range" which has not happened before.R stn :c+ 2-, 2.1 ma, 60 us, 130 hz battery 75%.Impedances: therapy impedance: 2195 ohm, 4.6 v 3- "not used", 2 - "avoid", 1- and 0- "investigate".Electrode impedances: monopolar (all outside normal range) 0: 6418, 1: 4351, 2: 3055, 3: 3055.Bipolar: abnormal impedances listed: 3-0: 4351, 2-0: 4375, 3-2: "low".An x-ray showed discontinuity in the lead about 1 1/2 cm above the connector.Will be turning device on and off and seeing how patient therapy is.The issue was not resolved at this time.
|
|
Search Alerts/Recalls
|
|
|