(b)(6).Method: the complaint rt380 evaqua2 adult breathing circuits were returned to fisher & paykel healthcare (f&p) in new zealand for evaluation, where it was visually inspected and analysed.Results: visual inspection of the returned circuits revealed that the expiratory tube of a rt380 adult dual heated evaqua2 breathing circuits were degraded.Ftir testing revealed that the tubing had been exposed to chemicals that induced hydrolysis and degradation.Conclusion: we are unable to determine the cause of the reported event.However, based on our investigation, it is likely due to chemical exposure to the expiratory tube.All rt380 adult evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject adult breathing circuits would have met the required specifications at the time of production.Our user instructions that accompany the rt380 adult evaqua2 breathing circuit state the following: - "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." - "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." - "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.".
|