Catalog Number 8708060
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/02/2022
Event Type  malfunction  
Event Description
It was reported that prior to a port placement procedure, hair was allegedly found near handle part of the sheath.The procedure was completed using another device.There was no patient contact.
 
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.(expiry date: 04/2023).
 
Event Description
It was reported that prior to a port placement procedure, hair was allegedly found near handle part of the sheath.The procedure was completed using another device.There was no patient contact.
 
Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one unsealed powerport titanium isp implantable port kit was returned for evaluation.Gross visual evaluation was performed.Although the sample was returned for evaluation, four electronic photos were provided for review.The investigation is confirmed for the reported contamination issue, as a piece of hair was noted to expose under the valve cap of the 8.0fr peel-apart sheath.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 04/2023), g3.H11: h6 (method, result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT ISP TI 8F CHRONWSH
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14825893
MDR Text Key298401570
Report Number3006260740-2022-02434
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741098895
UDI-Public(01)00801741098895
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K072215
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8708060
Device Lot NumberREFX4064
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-