Model Number 1012455-12
Device Problems Difficult to Remove (1528); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/24/2022
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue. The investigation determined the reported complaints appear to be related to operational context. There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that the procedure was performed to treat a lesion in the heavily calcified and tortuous 90% stenosed right coronary artery (rca). The 5. 0 x 12 mm nc trek balloon dilatation catheter (bdc) was advanced; however, resistance was noted and the bdc was unable to cross. Resistance was noted during removal, related to the vessel and lesion. There was no adverse patient effect and no clinically significant delay in procedure. The procedure was completed with a non-abbott stent. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNC TREK CORONARY DILATATION CATHETER
Type of DeviceCORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14837593
MDR Text Key295487594
Report Number2024168-2022-06934
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08717648152146
UDI-Public08717648152146
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K103153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation
Type of Report Initial
Report Date 06/27/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number1012455-12
Device Catalogue Number1012455-12
Device Lot Number00811G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/11/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-