MERIT MEDICAL SINGAPORE PTE LTD. SAFEDRAW¿; SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE Back to Search Results
Model Number 682000
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2022
Event Type  malfunction  
Event Description
The account alleges that during the hemodynamic monitoring of a patient, the pressure monitoring [pm] set was found to be leaking fluid from the 3-way stopcock.No patient injury to report.
 
Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Manufacturer Narrative
The suspect device has been returned for evaluation.The product was examined visually.The complaint is confirmed.No definitive root cause could be determined however, it is likely that significant force was applied to the device during use.A review of the device history and complaint database could not be performed since the lot number was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFEDRAW¿
Type of Device
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE
Manufacturer (Section D)
MERIT MEDICAL SINGAPORE PTE LTD.
198 yishun ave 7
singapore, sgp 76892 6
SN  768926
Manufacturer (Section G)
MERIT MEDICAL SINGAPORE PTE LTD.
198 yishun ave 7
singapore, sgp 76892 6
SN   768926
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
8012084662
MDR Report Key14861233
MDR Text Key299976648
Report Number8020616-2022-00044
Device Sequence Number1
Product Code DXN
UDI-Device Identifier00886333620005
UDI-Public00886333620005
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K935782
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number682000
Device Catalogue Number689439/JPB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received07/12/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-