Model Number G49100
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/24/2022
Event Type  malfunction  
Event Description
As reported, prior to use during a procedure involving endovascular repair of an abdominal aortic aneurysm, a flexor ansel guiding sheath's dilator would not flush.A photo provided by the customer shows an unknown substance within the lumen of the dilator tip.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Manufacturer Narrative
There is conflicting information from the customer regarding device return.(b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Manufacturer Narrative
Corrected information: d9, h3: the device was returned to cook for investigation.H3: device evaluated by mfg = other (81) - device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information has been received since the last report was submitted.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Corrected information: h6 (annex g).Summary of event: as reported, prior to use during a procedure involving endovascular repair of an abdominal aortic aneurysm, a flexor ansel guiding sheath's dilator would not flush.A photo provided by the customer shows an unknown substance within the lumen of the dilator tip.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation.The dilator was able to be flushed, but only when pressure was used.Foreign matter (possibly water) was noted at the distal tip of the device; however, after flushing the dilator, the foreign matter was no longer visible.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The product ifu instructs the user to flush the dilator with heparinized saline.The information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured out of specification; however, there is no evidence of additional non-conforming devices in-house or in the field.There are 100% inspections in place to capture this failure mode prior to distribution.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that manufacturing and quality control deficiencies contributed to this event.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information has been received since the last report was submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR ANSEL GUIDING SHEATH
Type of Device
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC.
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14876462
MDR Text Key303251608
Report Number1820334-2022-01137
Device Sequence Number1
Product Code DYB
UDI-Device Identifier00827002491007
UDI-Public(01)00827002491007(17)250328(10)14624541
Combination Product (y/n)N
PMA/PMN Number
K142829
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG49100
Device Catalogue NumberKCFW-8.0-18/38-45-RB-ANL1-HC
Device Lot Number14624541
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-