MEDTRONIC NAVIGATION, INC STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735665
Device Problems Material Integrity Problem (2978); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2022
Event Type  malfunction  
Event Description
Medtronic received information regarding a navigation system being used outside of a procedure. It was reported that the camera was not working, and had a constant orange led lit. There was no patient involvement.
 
Manufacturer Narrative
Other relevant device(s) are: product id: c amera 9735821 vega base s8 svc. A medtronic representative went to the site to test the equipment. Testing revealed that the camera and cable were replaced. The navigation system then passed the system checkout and was found to be fully functional. H6: b01, c02 and d02 apply to the system checkout. Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803. This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda. Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report. This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report. In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿. These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting. Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms. This statement should be included with any information or report disclosed to the public under the freedom of information act. Any required fields that are unpopulated are blank because the information is currently unknown or unavailable. A good faith effort will be made to obtain the applicable information relevant to the report. If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION¿ S8 SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14876834
MDR Text Key303145025
Report Number1723170-2022-01006
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 08/22/2022
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received08/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-