• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAZOR ROBOTICS LTD MAZOR X SYSTEM; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAZOR ROBOTICS LTD MAZOR X SYSTEM; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number TPL0059
Device Problems Imprecision (1307); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2022
Event Type  malfunction  
Manufacturer Narrative
Other relevant device(s) are: product id: asm0206, serial/lot #: sra00535, a medtronic representative went to the site to test the equipment.Testing revealed that the surgical arm failed the bist and stress tests.The surgical arm was recalibrated.The system then passed the system checkout and was found to be fully functional.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a guidance system being used during a spinal procedure.It was reported that there was a deviation during a single position lateral procedure.The patient was in the lateral position with their left side up.Registration was successfully completed with no shifting.Screws were placed on the lower side first and then moved to the top side.The surgical arm was sent to right l5 and the screw missed laterally by less than 1 cm.All of the other screws were accurate.The manufacturer representative noted that the surgical arm did fail a built in self test (bist).A hard reboot was done, but the surgical arm still failed the bist test.The deviated screw was removed and repositioned using a pak needle.There was no patient harm and the procedure was delayed less than an hour.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAZOR X SYSTEM
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IS  3079567
Manufacturer (Section G)
MAZOR ROBOTICS LTD
5 shacham street
p.o. box 3104
caesarea hefa,il 30795 67
IS   3079567
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14883882
MDR Text Key303343199
Report Number3005075696-2022-00052
Device Sequence Number1
Product Code OLO
UDI-Device Identifier07290109184838
UDI-Public07290109184838
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTPL0059
Device Catalogue NumberTPL0059
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age58 YR
Patient SexFemale
Patient Weight59 KG
-
-