• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG ENNOVATE SET SCREW STERILE; IMPLANTS POSTERIOR STABILISATI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG ENNOVATE SET SCREW STERILE; IMPLANTS POSTERIOR STABILISATI Back to Search Results
Model Number SY001TS
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problem Failure of Implant (1924)
Event Date 06/14/2022
Event Type  Injury  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with sy001ts - ennovate set screw sterile.According to the complaint description, there was a revision of a screw due to loosening with prospace xp cage.When locking with new nuts, the torque did not click a few times, but the nuts were spinning.A revision surgery was necessary.Additional information was not provided.Additional information has been requested but not yet received as of this report.Additional patient information is not available.The adverse event is filed under aag reference (b)(4).Associated medwatch-reports: 9610612-2022-00186 (400560279 - sy001ts).
 
Event Description
The adverse event is filed under aag reference (b)(4).Associated medwatch-reports: 9610612-2022-00183 (b)(4).9610612-2022-00186 (b)(4).
 
Manufacturer Narrative
Additional information: d9 - product not returned.Investigation: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level ((4) severity x (1) probability of occurrence) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigation results, a capa is not required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENNOVATE SET SCREW STERILE
Type of Device
IMPLANTS POSTERIOR STABILISATI
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key14917554
MDR Text Key295270094
Report Number9610612-2022-00183
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K180433
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSY001TS
Device Catalogue NumberSY001TS
Device Lot Number52734614
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/14/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-