HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
Model Number 1103 |
Device Problems
Pumping Problem (3016); Noise, Audible (3273)
|
Patient Problems
Thrombosis/Thrombus (4440); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/04/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient received a heart transplant because the ventricular assist device (vad) was exhibiting high watt alarms and grinding.The patient had no signs or symptoms prior to the transplant.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for corrections.B5 was corrected to indicate that a thrombus was suspected.In h6, the patient code was corrected from e2403 to e0514.Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that a thrombus was suspected.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for device evaluation.Product event summary: the pump ((b)(6)) was not returned for evaluation.Review of the controller log files revealed normal power consumption within the last 14 days of available data leading up to 04/jun/2022; no high watt alarms have been logged within the analyzed period.Additionally, one (1) low flow alarm was logged on 05/jun/2022 at 00:01:54.As result, the reported high power event was not confirmed.The reported "pump grinding" could not be confirmed due to insufficient evidence.Information provided by the site indicated that the patient received a heart transplant because the ventricular assist device (vad) due to high power, grinding, and suspected thrombus.The patient had no signs or symptoms prior to the transplant.Based on the available information, the device may have caused or contributed to the reported event.Based on the risk documentation, the ¿pump grinding" event may be attributed to multiple factors including, but not limited to, thrombus within the device leading to impeller imbalance or the placement of the pump which allows contact with fixed or rigid anatomical structure.Based on the risk documentation, possible causes of the observed low flow event may be attributed to multiple factors including but not limited to thrombus at the inflow cannula/outflow graft, constriction at the outflow graft, and/or poor vad filling.Per the instructions for use, device thrombus is a known potential complication associated with the implantation of a vad.Based on review of past adverse events for this patient, it was noted that the patient had a history of thrombus.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|