It was reported that due to the patient's small left ventricle and body size, the position of the pump cannula was close to the lateral wall.A surgical correction had already been performed weeks ago by anchoring the pump with a ligature through the pericardium.However, the patient still continued to experience intermittent low flow alarms that often occurred when the patient moved.The site stated no further re-positioning of the pump was possible.The patient was noted to be clinically stable.A low flow 2.0 software upgrade was successfully performed and the low flows alarms were reduced.
|
Manufacturer's investigation conclusion: the reported low flow alarms could not be confirmed as no log files were provided by the account for review.The account reported that the alarms were due to the patient¿s small body size and left ventricle.A direct correlation between heartmate (hm) 3 left ventricular assist system (lvas), serial number (b)(6), and the reported events could not be conclusively determined.The patient remains ongoing on heartmate 3 lvas, serial number (b)(6) and no further related events have been reported at this time.The relevant sections of the device history records for (b)(6) were reviewed and showed no deviations from manufacturing or quality assurance specifications.The heartmate 3 lvas instructions for use (ifu) (rev.G) and the heartmate 3 patient handbook (rev.G) are currently available.Section 1 of the ifu, ¿introduction¿, lists potential adverse events that may be associated with the use of the heartmate 3 left ventricular assist system.Section 1 of the ifu also addresses all pump parameters, including pump flow.Section 4 of the ifu, ¿system monitor¿, describes the pump flow display and the hazard alarms.Per design, when the estimated flow value is calculated at less than 2.5 lpm, a low flow status is posted to the log file.If the flow remains below 2.5 lpm for 10 seconds, a low flow hazard alarm is triggered.Section 4 also explains that changes in patient condition can result in low flow.Section 7 of the ifu, ¿alarms and troubleshooting¿, and section 5 of the patient handbook, ¿alarms and troubleshooting¿, address all system alarm conditions as well as the appropriate actions associated with each condition.Furthermore, section 8 of the patient handbook, ¿handling emergencies¿, also provides examples of emergencies and the proper actions to take in the event an emergency occurs.No further information was provided.The manufacturer is closing the file on this event.
|