Catalog Number SGC0702 |
Device Problems
Break (1069); Positioning Failure (1158); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is filed to report cable break.It was reported that the mitraclip procedure was performed on (b)(6) 2022 to treat functional mitral regurgitation (mr) with a grade of 4.When turning the + knob, a noise was heard.The "+" cable broke when the clip was oriented toward the valve.The steerable guide catheter (sgc) was unable to curve.The clip and the guide were removed together without difficulty.One clip was implanted and the mr was reduced to 1.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation was unable to determine a cause for the reported cable break.The positioning failure of inability to curve and noise were cascading effects of the broken cable.There is no indication of a product issue with respect to manufacture, design, or labeling.
|
|