Catalog Number 2C8634
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that an unspecified number of clearlink extension sets ruptured during attempted power injection during a computed tomography (ct) scan.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Initial reporter facility name: (b)(6) hospital for children.Device manufacturer address 1: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the devices were not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: baxter was unable to determine the cause for this report by sample evaluation, however, the probable cause is user related.The clearlink extension sets, involved in this report, have only been tested up to 45 psi and therefore, are not intended for use with a power injector.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK LUER ACTIVATED VALVE
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14946222
MDR Text Key296055892
Report Number1416980-2022-03514
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112893
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C8634
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
POWER INJECTOR
-
-