CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE Back to Search Results
Model Number MCH-1000 (I)
Device Problems Melted (1385); Device Emits Odor (1425); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/03/2019
Event Type  malfunction  
Event Description
Customer complains device was in defrost mode in preparation to clean and burning was smelled by perfusionists.
 
Manufacturer Narrative
The manufacturer is reporting the following complaint after a voluntary review of all complaints (reportable or not) since 2016.This report is being filed now, after being scrutinized under a newly revised risk matrix, recently adopted after inspection.Upon investigation, cardioquip found that the device plug and the hospital outlet had melted together.The hospital outlet that the device was connected to was inaccessible to the technician therefore, it could not be determined if it was the device plug or the hospital outlet that had caused the failure.The device plug was replaced and inspected for full functionality.
 
Event Description
Customer complains device was in defrost mode in preparation to clean and burning was smelled by perfusionists.
 
Manufacturer Narrative
Further review determined that this is a non-reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOQUIP MODULAR COOLER HEATER
Type of Device
CARDIOPULMONARY BYPASS DEVICE
Manufacturer (Section D)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer (Section G)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer Contact
charley ford
8422 calibration ct.
college station, TX 77845
9796910202
MDR Report Key14950036
MDR Text Key303716108
Report Number3007899424-2022-00046
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102147
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMCH-1000 (I)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-