• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES CELL-DYN RUBY SYSTEM; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES CELL-DYN RUBY SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number 08H67-01
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2022
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
An abbott field service engineer observed smoke coming from the cell-dyn ruby analyzer.There was no report of fire or any injury.
 
Manufacturer Narrative
The cell-dyn ruby instrument was in the daily maintenance status when the smoke was observed.There was no injury as no one was in the lab at that time.No fire was seen only smoke.No damages to the surrounding areas were reported.The electrician was called to check the facility and the instrument was unplugged.It was found that fluid from the instrument¿s waste drain was the source of the problem.There were some changes in the lab and the waste drain was not properly installed by the customer.After the waste drain was properly installed, the issue was resolved.A review of all tickets associated with the cell-dyn ruby and a review for complaint trends for the list number was performed.The review did not identify any adverse trends and the complaint activity was normal.Additionally, labeling was reviewed and adequately addressed the issue.Based on this information, a deficiency was not identified.A malfunction was not identified.Based on the information within the complaint record, the device met performance specifications and performed as intended at the customer site.Use error caused the issue as the waste drain was not properly installed by the user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN RUBY SYSTEM
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
ABBOTT LABORATORIES
4551 great america parkway
santa clara CA 95054
Manufacturer (Section G)
ABBOTT LABORATORIES
4551 great america parkway
santa clara CA 95054
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14961429
MDR Text Key303098062
Report Number2919069-2022-00019
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740017170
UDI-Public00380740017170
Combination Product (y/n)N
Reporter Country CodeBD
PMA/PMN Number
K061667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number08H67-01
Device Catalogue Number08H67-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/20/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-