Model Number 250
Event Date 07/29/2009
Event Type  Malfunction  
Event Description

It was reported that a neurologist's handheld was showing multiple times of magnet activations. The neurologist called the manufacturer for further instructions. Further info from the manufacturer to the neurologist indicated that the pt's generator was fine and that it was related to roll-over of total time and magnet activations which should resolve once the pt has had fifteen magnet activations. The neurologist acknowledged and he stated that he would re-interrogate the pt's generator at the next visit. Moreover, the neurologist stated the pt was doing well with vns therapy and had no problems. Furthermore, good faith attempts to obtain additional info have been unsuccessful to date.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS, INC.
houston TX 77058
Manufacturer Contact
nydia grimes
100 cyberonics blvd., ste. 600
houston , TX 77058
2812287200
MDR Report Key1497017
Report Number1644487-2009-01850
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 07/29/2009
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/28/2009
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number250
OTHER Device ID NumberVERSION 7.1
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Date Manufacturer Received07/29/2009
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured10/01/2008
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

-
-