This report is part of a retrospective review and remediation efforts in response to a warning letter.(b)(4).S/w 2.9d the unit did not have a battery installed when received.Unit passed the functional test, including the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and the dat test at 0.08730 inches.Unit uploaded properly using carelink.The formatted history file lists multiple suspend activity.The latest suspend activity was on (b)(6) 2017 for low predicted suspend, suspend predict threshold, threshold suspend input, tsuspendmxst, suspended by alarm, and low threshold suspend.The long trace history file lists some data on (b)(6) 2017 for button presses activity.The suspended and resume options functions properly when using the suspend delivery feature.No anomaly noted when using the suspend delivery feature.Unit had scratched case, pillowing keypad overlay and cracked retainer.The test p-cap and reservoir does lock in place in the reservoir compartment.Data analysis: from 10/04/2017 to 10/10/2017 daily total of all insulin delivered = 0.On (b)(6) 2020: the case bottom was milled off to perform a visual inspection.During the visual inspection, no moisture damage or loose components noted on the electronic assembly.No damage noted on the motor assembly.
|