Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2022
Event Type  malfunction  
Event Description
It was reported during the incoming inspection at the warehouse in japan, a team member found debris in the sterile package.No adverse events have been reported as a result of this malfunction.Additional information on the reported event is unavailable.
 
Manufacturer Narrative
(b)(4).Foreign country: japan.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 -2022 -01600.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: h2; h3; h6.Evaluation of the returned product confirmed foreign debris is present inside the sterile packaging, and the sterile packaging remains sealed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the provided work instructions during manufacturing.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 FP TYPE1 PPS HO 5.0
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14985972
MDR Text Key300443154
Report Number0001825034-2022-01599
Device Sequence Number1
Product Code KWL
UDI-Device Identifier00880304499065
UDI-Public(01)00880304499065(17)311020(10)7124293
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K103755
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-101050
Device Lot Number7124293
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2022
Was the Report Sent to FDA? No
Date Manufacturer Received08/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-