ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752192
Device Problems Device Alarm System (1012); Inability to Irrigate (1337)
Patient Problems Corneal Edema (1791); Eye Burn (2523)
Event Date 06/27/2022
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported inadequate fluid was irrigating the eye and the patient suffered a wound burn.The procedure was completed after the product was replaced.The patient required suturing of the wound.The sample is not available for return as it was discarded.Additional information received from the company representative who indicated occlusion tone was heard during the procedure.The corneal burn resulted in mild gaping at end of procedure but no leak was observed with wound hydration.The patient was sutured with one nylon 10/0 for safety.The injury resulted in postoperative astigmatism which valued to be d1 6/12 (unaided) and further review for assessing astigmatism was planned in a week.The burn resulted in corneal opacity which was cleared significantly after the epithelium was cleared off.The patient was stable on day-1 post operation.Additional information received from the company representative who indicated the patient's current condition is improving.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.No sample has been returned for evaluation; therefore, the condition of the product could not be verified.Visual inspection or functional testing could not be conducted in order to ascertain the failure mode for the consumable device.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.Based on our current tracking, there are no adverse trends for this reported complaint.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14999483
MDR Text Key295772385
Report Number1644019-2022-00558
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657521920
UDI-Public00380657521920
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number8065752192
Device Lot Number2532010H
Was Device Available for Evaluation? No
Date Manufacturer Received09/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
-
-