• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number 24630
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/24/2022
Event Type  malfunction  
Event Description
It was reported that a balloon rupture occurred.The 75% stenosed target lesion was located in the moderately calcified left cephalic vein.A 5.00mm/2.0cm/90cm peripheral cutting balloon was selected for use.During the procedure, at first inflation, it was noted that the balloon ruptured in a pinhole form at the middle part at 6atm within 5 seconds.The device was removed using normal method without any problem.According to the physician, it was probable that the device ruptured because the lesion part was stiff.The procedure was completed with another of same device.No complications were reported, and patient was good post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2CM PERIPHERAL CUTTING BALLOON
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15003686
MDR Text Key295877449
Report Number2124215-2022-25492
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/15/2023
Device Model Number24630
Device Catalogue Number24630
Device Lot Number0028561195
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GUIDEWIRE - 0.018INCH FUGA; INFLATION DEVICE - EVEREST; INTRODUCER SHEATH - 6F MEDIKIT SHEATH
-
-