Model Number 106524US
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abdominal Pain (1685); Bacterial Infection (1735); Purulent Discharge (1812); Skin Infection (4544)
Event Date 06/17/2022
Event Type  Injury  
Event Description
It was reported that the patient was admitted from the clinic on (b)(6) 2022 due to purulent drainage and abdominal pain.The patient underwent a driveline debridement on (b)(6) 2022.The wound cultures came back positive for corynebacterium.Patient was discharged home on (b)(6) 2022 on iv vancomycin 1g q 12hrs.Antibiotic stop date was set to be (b)(6) 2022.No other patient issues or concerns were noted during admission.The pump reportedly functioned appropriately.
 
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the pump remains in use supporting the patient.A correlation between the device and the report of infection could not be conclusively determined.Infection is listed in the instructions for use as a potential adverse event that may be associated with the use of heartmate 3 left ventricular assist system.Infection has been previously investigated and will continue to be monitored through quality data reviews, which are conducted on production and post product signals to evaluate if products are conforming to product requirements.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 LVAS IMPLANT KIT
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer (Section G)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key15005510
MDR Text Key295856958
Report Number2916596-2022-12001
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00813024013297
UDI-Public00813024013297
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/13/2023
Device Model Number106524US
Device Catalogue Number106524US
Device Lot Number7823822
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/30/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age24 YR
Patient SexMale
Patient Weight63 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-