The affected device was tested and repaired on site by a field service technician - a printed circuit-board ("vgc analog 2" for processing of analog signals and actors and the encoder signals for ventilation control) has been exchanged and was sent together with the saved device log file to the manufacturer for investigation.The analysis of the device log file confirmed a problem with calibration data (pressure sensor) which has been detected during start up of the device, i.E.During pre-use check in the morning (without patient).According the log entries the issue was solved by calibration procedure (initiated by the user) and the device went into service, i.E.Use on a patient - the following log entries are unconspicous.The log entries does not confirm the initially reported ventilator fail during use on a patient.The exchanged pcb was tested for 7 days including multiple pre-use checks - the reported symptom could not be reproduced.Thus, the exact root cause for initially reported ventilator fail during use could not be determined.H3 other text: device checked and repaired on-site, log file and exchanged printed circuit board returned to manufacturer for investigation, exact root cause could not be determined.
|