The hospitalâs anesthesia technician and ge field engineer found no evidence of device malfunction, and the device was returned to service after testing.The root cause of the patient o2 desaturating when the ventilator was switched to manual ventilation cannot be determined.It is possible that a combination of the adjustable pressure limiting valve being set too low, leaks or obstructions within the patient breathing circuit accessories, tubing connections not fully seated, kinked tubing, leaks within the bag, customer forgetting to manually bag when switched to manual mode, or providing less minute volume within manual ventilation as previously delivered during mechanical ventilation, may have contributed to the o2 desaturation.
|