Catalog Number H938724
Device Problems Filling Problem (1233); Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/22/2022
Event Type  malfunction  
Event Description
It was reported that during compounding with two (2) exactamix 2400 valve sets, the volume of the final product was greater than 3% off.The technician called baxter customer service, who had the technician switch to another port, but this did not resolve the issue.Upon follow-up, the customer reported that they noticed air on port 22 where dextrose 70% was hanging.There was no damage observed to the valve sets.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
One (1) device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H4: the lot was manufactured between april 07, 2022 - april 12, 2022.H10: one (1) actual device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.System level testing was performed and a leak or volume issue was not observed.The sample was evaluated by attaching a solution to ports 2 and 22 and then analyzed at various points throughout the prime and verify process and pump calibration process.No leaks or bubbles were observed.The reported conditions were not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15029487
MDR Text Key297085848
Report Number1416980-2022-03677
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477183
UDI-Public(01)00085412477183
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K002705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938724
Device Lot Number60359323
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received07/18/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-