MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP/DRIVELINE VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP/DRIVELINE VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1103
Device Problems Electrical /Electronic Property Problem (1198); Pumping Stopped (1503); Material Integrity Problem (2978); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2022
Event Type  Injury  
Event Description
It was reported that the ventricular assist device (vad) patient presented after the controller exhibited an electrical fault alarm and several vad stop alarms. There were no further alarms, and the connector and driveline were manipulated without being able to replicate the alarm. It was also noted that there was a cut in the driveline sheath. The patient was admitted and a driveline sheath repair was performed. The vad/driveline and the controller remain in use. No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Additional products: brand name: heartware ventricular assist system ¿controller 2. 0, model #: 1420/ catalog #: 1420/ expiration date: 28-feb-2022 / serial or lot#: (b)(4), udi #: (b)(4), device available for evaluation: no, device evaluated by mfr: no, device evaluation anticipated, but not yet begun, mfg date: 18-feb-2021, labeled for single use: no, (b)(4). Investigation of this event is pending and a supplemental report will be sent upon its completion. Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803. This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda. Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report. This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report. In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿. These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting. Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms. This statement should be included with any information or report disclosed to the public under the freedom of information act. Any required fields that are unpopulated are blank because the information is currently unknown or unavailable. A good faith effort will be made to obtain the applicable information relevant to the report. If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP/DRIVELINE
Type of DeviceVENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key15052819
MDR Text Key296143992
Report Number3007042319-2022-06633
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/19/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/31/2022
Device Model Number1103
Device Catalogue Number1103
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/13/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 07/19/2022 Patient Sequence Number: 1
-
-