HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
Model Number 1420 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/20/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the shoulder pack was unable to be used because the controller would get too hot.The shoulder pack was removed from use.The controller remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Additional products: heartware ventricular assist system ¿ shoulder pack model #: 2060 / catalog #: 2060 / expiration date: 31-oct-2022 / lot #: 1787915 udi #: (b)(4).Device available for evaluation: no, mfg date: 18-oct-2021, labeled for single use: no (b)(4).Additional information has been requested regarding the details of the event, but it was not available at the time of this report.If additional information is received, the event will be updated and a supplemental report will be sent.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for investigation completion.Product event summary: the controller (b)(6) and the shoulder pack (lot 1787915) were not returned for evaluation.As a result, the reported event could not be confirmed.Applicable risk documentation and experience with events of similar circumstances were considered; a possible root cause of the reported event may be attributed, but not limited, to environmental factors and/or a controller malfunction.Additional products: d4: lot #: 1787915 h6: fda method code(s): b17 h6: fda results code(s): c20 h6: fda conclusion code(s): d14 investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|