Model Number 0502-404-030
Device Problems Crack (1135); Poor Quality Image (1408)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2022
Event Type  malfunction  
Event Description
It was reported that there was blurry image during procedure.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Manufacturer Narrative
This scope was not received for evaluation at stryker endoscopy.This scope was received at henke for evaluation.Based on the henke service record attached, the reported failure "cracked lens" was confirmed.According to henke: scope evaluated as a level 3.Distal tip and fiber damage, cracked on distal lens negative.Probably root cause for this failure is: the complaint for ¿cracked lens¿ has been confirmed and is due to customer use and handling.The reported failure mode will be monitored for future reoccurrence.The device manufacturer date is not known.
 
Event Description
It was reported that there was blurry image during procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION IE 4K C-MOUNT ARTHROSCOPE, AUTOCLAVABLE, 4MM X 30°, 140MM, SPEED-LOCK
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15083612
MDR Text Key296585618
Report Number0002936485-2022-00422
Device Sequence Number1
Product Code HRX
UDI-Device Identifier07613327416497
UDI-Public07613327416497
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K183470
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0502-404-030
Device Catalogue Number0502404030
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received06/28/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse