• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP SHELL HA W. FLANGES 51 NON-CEM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP SHELL HA W. FLANGES 51 NON-CEM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED Back to Search Results
Catalog Number 75017242
Device Problem Loss of Osseointegration (2408)
Patient Problems Failure of Implant (1924); Pain (1994); Inadequate Osseointegration (2646)
Event Date 08/26/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that, after a right tha had been performed on (b)(6) 2008, the patient experienced pain, loosening of the acetabular cup and wear of the liner.A revision surgery was performed on (b)(6) 2021 to treat this adverse event.During this procedure, the acetabular cup, the liner and head were explanted.The patient's health status is not known.This information was provided by the national joint registry of the united kingdom, as part of a retrospective data collection of patients who underwent either a revision or re-revision surgery where a polarcup shell was part of the hip construct.As such, no further information will be available.
 
Manufacturer Narrative
A total of 70 revision surgeries involving polarcup devices was reported by the uk national joint registry.104 complaints were opened accordingly.A review of the device history record (dhr), of the complaint history (chr), of the instruction for use (ifu), of the risk management file (rmr) and of past escalation (per) was performed.No relevant supporting clinical information was provided.Based on insufficient information, a thorough clinical assessment cannot be performed at this time.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined.Based on the chr and dhr evaluation performed there is no indication that any failure to meet specification contributed to the reported events.The reported failure modes are outlined in the device instruction for use lit no 12.23 03/21.The occurrence and severity of the reported issues are in line with the risk plans.The root cause of the reported events was attributed to known inherent risks of the devices.No corrective actions have been initiated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP SHELL HA W. FLANGES 51 NON-CEM
Type of Device
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15086307
MDR Text Key296410483
Report Number9613369-2022-00416
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K122244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Study,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/01/2008
Device Catalogue Number75017242
Device Lot NumberA0802439
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received07/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
-
-