Model Number PM3562
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fever (1858); Unspecified Infection (1930)
Event Date 06/29/2022
Event Type  Injury  
Manufacturer Narrative
A device history record (dhr) review was performed and review of the sterilization records confirmed normal sterilization cycles for the products.The device met specifications prior to leaving abbott manufacturing facilities.The product was returned and visual inspection was normal.The cause of infection could not be traced to the device. .
 
Event Description
Related manufacturer reference number: 2017865-2022-15885, related manufacturer reference number: 2017865-2022-15886, related manufacturer reference number: 2017865-2022-15887.It was reported the patient presented to the emergency room with fever, chills, redness with excess fluid in the pocket and septic shock.Blood cultures taken from the patient tested positive for meticillin-sensitive staphylococcus aureus (mssa) bacteremia.The entire pacemaker system was explanted.The patient was recovering after the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15089707
MDR Text Key296456231
Report Number2017865-2022-15884
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberP000132753
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention; Hospitalization;
Patient Age85 YR
Patient SexMale
-
-