Catalog Number ROB00036
Device Problem Interrogation Problem (4017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/04/2022
Event Type  Injury  
Event Description
It was reported that, during a navio assisted surgery, during femur cut stage of the surgery, the error e8 came on the anspach console.The procedure was completed with a delay of less than 30 minutes, with a change of technique to a manual procedure.The patient was not harmed beyond the reported problem.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
H3, h6.The navio surgical system india, part number rob00036, serial number (b)(6), used for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.The clinical/medical investigation concluded that the device did not perform as intended during the ¿femur cut stage¿ of the surgery, and the navio system was abandoned.Per the field report form, the surgery was concluded using manual procedure with a 0-30 minute surgical extension without patient injury due to the reported event.Patient impact beyond the reported 0-30 minute surgical delay, abandoning the navio¿ system and subsequent use of manual instrumentation to perform the procedure would not be anticipated as manual procedure/standardized instrumentation is an approved surgical technique and no patient harm was alleged due to the reported event.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.If a navio surgical system failure occurs at any point during the surgical case, refer to the surgical technique guide for knee arthroplasty, recovery procedure guidelines tables for proper guidance regarding recovering to a fully manual procedure.A failure can consist of, but is not limited to, a system software crash, unrecoverable hardware failure, handpiece failure with no backup available, tracker array failure or loss of contact with bone that is unrecoverable, etc.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still within the product risk profile.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with a drill/handpiece issue.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO SURGICAL SYSTEM INDIA
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15090624
MDR Text Key296460470
Report Number3010266064-2022-00538
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556693506
UDI-Public885556693506
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberROB00036
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/06/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKN NAVIO ROBOTICS INSTR
Patient Outcome(s) Required Intervention;
-
-