Model Number 97715
Device Problem Overheating of Device (1437)
Patient Problems Burn(s) (1757); Erythema (1840); Pain (1994); Peeling (1999); Fluid Discharge (2686); Alteration in Body Temperature (4568)
Event Date 06/22/2022
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient via healthcare provider who was implanted with an implantable neurostimulator (ins) for spinal pain indications.The caller reported the patient has redness at implant site, that it looked like a burn circle around the battery, patient has pain and it's hot to touch, with drainage and skin peeling off.Issue started last week.Technical services(ts) reviewed with caller that recharging shouldn't cause this since there is a limit to how much the system heats up before recharging will stop.Redirected caller to hcp to assess.Troubleshooting was not required.The issue was not resolved through troubleshooting.The patient was redirected to their healthcare provider to further address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key15098916
MDR Text Key304554348
Report Number3004209178-2022-09419
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00763000315467
UDI-Public00763000315467
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Date Manufacturer Received06/29/2022
Date Device Manufactured03/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR
Patient SexMale
-
-