• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. MOVATION KNEE DOMED TRI-PEG PATELLA, 35X9MM, E-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. MOVATION KNEE DOMED TRI-PEG PATELLA, 35X9MM, E-PLUS Back to Search Results
Model Number 130-03-735
Device Problems Break (1069); Appropriate Term/Code Not Available (3191)
Patient Problem Fall (1848)
Event Date 07/13/2022
Event Type  Injury  
Event Description
Revision surgery - patient fell, broke patellar component, revised patella and put a new liner in.
 
Manufacturer Narrative
The reason for this revision surgery was reported as fall. The previous surgery and the surgery detailed in this event occurred 2 years and 2 months apart. The healthcare professional indicated there was no delay in surgery and another suitable device was available for use. The revision surgery was completed as intended. The devices were disposed of at hospital and not made available to djo surgical for examination. A review of the device history records (dhr) show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements. There was a nonconformance associated with the main part #130-03-735, domed tri-peg patella, 35x9mm, e-plus which documents that out of 20 parts lot, 1 part was rejected and scrapped due to surface finish not acceptable. All other items in the lot were met with the design, fit and function requirements. The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery. Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review. The root cause of this complaint was a revision surgery due to fall. There were no findings during this evaluation that indicate the reported devices were defective. No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event. Agent has clearly mentioned that "patient fell" and there are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities. There are no indications of a product or process issue affecting implant safety or effectiveness. Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOVATION KNEE
Type of DeviceDOMED TRI-PEG PATELLA, 35X9MM, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key15099506
MDR Text Key296558092
Report Number1644408-2022-00991
Device Sequence Number1
Product Code OIY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model Number130-03-735
Device Catalogue Number130-03-735
Device Lot Number156N1591
Was Device Available for Evaluation? No
Date Manufacturer Received07/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/31/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 07/25/2022 Patient Sequence Number: 1
Treatment
342-16-708 LOT 162T1025
-
-