LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number 16-02-80
Device Problem Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2022
Event Type  malfunction  
Event Description
Livanova deutschland received a report that the residual current circuit breaker triggers when a heater-cooler system 3t device is switched on.There was no patient involvement.
 
Manufacturer Narrative
There was no known patient involvement.The heater-cooler (b)(4) 80 is not distributed in the usa, but it is similar to heater-cooler (b)(4) 85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in koeln, germany.The fault current is triggered by a defective compressor.The device should be discarded and disposed of.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: the reported event is a known issue.Livanova put in place corrective action to reduce this type of events in case of breach of the cooling coil being generated due to stirrer flow in the tank leading to hole formation in a narrow area of the coil.However, in this specific case, the exact location of the hole on the cooling coil could not be determined.In october 2019 it was performed the erosion kit upgrade of the device that includes the new design of the pump head of the stirrer motor to prevent the water flow directed to the narrow area of the cooling coil.The reported event was claimed after the upgrade and it is not possible to exclude that the condition of the cooling coil was already compromised before the upgrade to such an extent that the copper was rather degraded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key15101356
MDR Text Key296563855
Report Number9611109-2022-00362
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number16-02-80
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/30/2022
Date Device Manufactured10/17/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse