CONVATEC DOMINICAN REPUBLIC INC L3W0750 - DUODERM; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Model Number 187662
Device Problems Delivered as Unsterile Product (1421); Packaging Problem (3007)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that there was seal on dressing reported during inspection.The product was not used.The photographs depicting the issue were received from the complainant.
 
Manufacturer Narrative
Device 2 of 4.Reporter affiliation - (b)(6).Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: h6 - investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions h10: investigation summary batch record review: the lot 1f03164 was manufactured on 23 june 2021 uhlmann#2 manufacturing line, with a total of(b)(4) market units.Complaint performed a batch record review on 30 october 2022, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct.Sap (system application product) material 1704763 and manufacturing order (b)(4).The batch record review supports that there were no discrepancies related to the issue reported.Returned sample evaluation: photo related to the reported problem is available for evaluation.Conclusion summary of the related event: the scope of this investigation covers the failure mode related to the malfunction code primary pack has incomplete or open seal / weld, or is torn, ripped or contains holes or exhibits external contamination, or dressing or loose material is trapped in packaging for doyen b and uhlmann #2 line, located at awc dominican republic facility.Where all products with control gel formula cgf and extra thin dressing are impacted.The risk level is classified as moderate based on the risk evaluation performed as part of the scope.The po rate of 1 (remote) was determined, which means that is highly unlikely that the harm will occur, which correspond to the reported complaints, as no harm to the end user happened.Based on the severity of the failure mode, categorized as critical on the event, a stop shipment was initiated for the impacted lots.In order to determine the possible causes of the malfunction code primary pack has incomplete or open seal / weld, or is torn, ripped or contains holes or exhibits external contamination, or dressing or loose material is trapped in packaging for doyen b and uhlmann #2 line, a brainstorming tool was used to determine all the possible causes to finally summarize them in a cause & effect diagram to regroup the finding according to six (6) m¿s (materials, process/method, measurement, manpower, machinery/equipment and environment) and discard or retain the potential causes.Based on the investigation findings, the following root causes were found per line and failure mode reported: doyen b caught in seal: -method, dressing not into infeed conveyor pitch -method, manual setting of indexation process during changeover -method, tensor of material positioning -machine, incorrect dressing tooling on machine uhlmann #2, caught in seal: - method, lack of visual help for manual dressing placement: - machine, suction cup worn out from pick and place: a capa (corrective and preventive actions) plan will be generated to track the implementation of this actions and measure the effectiveness in the product and the process.This issue will be monitored through the post market product monitoring review process, standard operating procedure (sop).To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Event Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L3W0750 - DUODERM
Type of Device
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR  91000
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR   91000
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 
3365424679
MDR Report Key15101535
MDR Text Key296861331
Report Number9618003-2022-04233
Device Sequence Number1
Product Code NAD
UDI-Device Identifier00768455174928
UDI-Public00768455174928
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number187662
Device Lot Number1F03164
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received11/01/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/23/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-