Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
A supplemental report is being submitted for device analysis; product event summary: the battery (b)(6) was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the devices in relation to the reported event.Log file analysis was not performed since log files were not available for analysis.Failure analysis of the returned device revealed that the unit passed visual inspection and functional testing.Of note, as received functional testing revealed that the battery had cell under voltage (cuv) and pack under voltage (puv) flags enabled, which lead to the discharge fet disabled.However, the battery was able to adequately charge during bench testing.After allowing the battery to charge the cuv and puv flags were reset, and the battery performed as intended.The battery was able to adequately connect to and provide power to a test controller, as well as charge and discharge with no anomalies observed.As a result, the reported event could not be confirmed.Based on the available information, a possible root cause of the reported "not charging" event can be attributed, but not limited to a marginal connection between the battery and battery charger.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|