Model Number MMT-712WWB
Device Problem Moisture or Humidity Problem (2986)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/25/2021
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the insulin pump was exposed to moisture.No harm requiring medical intervention was reported.The insulin pump was returned for analysis.
 
Manufacturer Narrative
(b)(4).Pump passed the displacement test and self test.However, moisture damage at electronic assembly.Also, moisture damage motor during visual inspection.Perform brush cleaning with the isopropyl alcohol and cleaning up all corrosion on the electrical board a and powered the pump on using the battery simulator and the unit powered up normally or no blank display noted.The following were noted during visual inspection: cracked belt clip slot, cracked battery tube thread and cracked reservoir tube lip.The test p-cap/reservoir does lock into place.After inserting battery unit power properly.Pump monitored for several days and no blank display noted.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MMT-712 PARADIGM INSULIN PUMP
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
7635140379
MDR Report Key15119089
MDR Text Key304586090
Report Number2032227-2022-303372
Device Sequence Number1
Product Code LZG
UDI-Device Identifier000000613994470898
UDI-Public(01)000000613994470898
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-712WWB
Device Catalogue NumberMMT-712WWB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2022
Date Manufacturer Received07/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/23/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-