On the event date (b)(6) 2021.Customer returned pump for an alleged cracked battery tube.Unable to perform displacement test and selftest due to constant pump error e52 alarm.Constant pump error e52 alarm due to suspected motherboard.Insulin pump history download using thds was successful.However, there is no found data available date & time of alarm on pump history file on the event date.Device was cut/open and inspected no moisture damage at the electronic assembly.Test reservoir lock properly inside the reservoir tube.Device perform visual inspection and pump received with cracked reservoir tube lip and broken battery tube threads.Confirmed pump error e52 alarm due to suspected at the m/b.Confirmed cracked reservoir tube lip and broken battery tube threads.(b)(4).Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|