• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1650
Device Problem Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.This information was received from the mechanical circulatory support product surveillance registry study.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported by the patient that one battery was no longer charging to full when being connected to the battery charger over long periods of time.The battery was replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
A supplemental report is being submitted for investigation completion.Product event summary: the battery was not returned for evaluation.Review of the manufacturing documentation confirmed that the associated device met all requirements for release.As a result, the reported event could not be confirmed.Applicable risk documentation and experience with events of similar circumstances were considered; events involving a battery not charging can be attributed, but not limited, to an internal battery communication error, faulty integrated circuit, improper weld, soldering defect, out of temperature range, and/or a charge time-out of eight (8) hours.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key15130118
MDR Text Key304182525
Report Number3007042319-2022-06751
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00888707006873
UDI-Public00888707006873
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/31/2021
Device Model Number1650
Device Catalogue Number1650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/27/2022
Date Device Manufactured03/21/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
1103 VAD
Patient Age69 YR
Patient SexMale
Patient Weight80 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-