C.R. BARD, INC. (BASD) -3006260740 POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Model Number 1606062 |
Device Problem
Difficult or Delayed Separation (4044)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/29/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
As the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device pending return.
|
|
Event Description
|
It was reported that during a port placement procedure, the sheath was allegedly unable to be peeled away well.The sheath was cut with scissors and used.There was no reported patient injury.
|
|
Event Description
|
It was reported that during a port placement procedure, the sheath was allegedly unable to be peeled away well.It was further reported that the sheath was cut with scissors and used.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one fully peeled 6.5fr peel-apart sheath was returned for evaluation.Gross visual evaluation was performed.Bends were noted throughout each fully peeled-apart sheaths.One peeled sheath shaft was noted to have a not fully detached valve.Both peeled sheath shafts were noted to be bunched.The investigation is inconclusive for the reported difficult to peel away sheath issue, as the exact circumstances at the time of the reported event cannot be verified, and the reported event could not be reproduced in the lab.A definitive root cause for the reported failure could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: g3 h11: h6 (method, result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|