• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 1606062
Device Problem Difficult or Delayed Separation (4044)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2022
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Device pending return.
 
Event Description
It was reported that during a port placement procedure, the sheath was allegedly unable to be peeled away well.The sheath was cut with scissors and used.There was no reported patient injury.
 
Event Description
It was reported that during a port placement procedure, the sheath was allegedly unable to be peeled away well.It was further reported that the sheath was cut with scissors and used.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one fully peeled 6.5fr peel-apart sheath was returned for evaluation.Gross visual evaluation was performed.Bends were noted throughout each fully peeled-apart sheaths.One peeled sheath shaft was noted to have a not fully detached valve.Both peeled sheath shafts were noted to be bunched.The investigation is inconclusive for the reported difficult to peel away sheath issue, as the exact circumstances at the time of the reported event cannot be verified, and the reported event could not be reproduced in the lab.A definitive root cause for the reported failure could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: g3 h11: h6 (method, result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT CLEARVUE ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 6F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15136703
MDR Text Key304012269
Report Number3006260740-2022-02986
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741026416
UDI-Public(01)00801741026416
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K063377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1606062
Device Catalogue Number1606062
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/17/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-