• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INGENIO; IMPLANTABLE PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INGENIO; IMPLANTABLE PACEMAKER Back to Search Results
Model Number K174
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems Fall (1848); Dizziness (2194); Presyncope (4410); Unspecified Heart Problem (4454)
Event Date 07/09/2022
Event Type  Injury  
Manufacturer Narrative
This investigation will be updated should further pertinent information be provided.
 
Event Description
It was reported that this pacemaker was explanted as it transitioned into safety mode.The patient felt palpitations, light headedness and fell once due to dizziness.A new non bsc device was implanted at the same surgical procedure.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode due to system resets.It was confirmed that brady therapy remained available.The system resets occurred during a telemetry session and caused the device to enter safety mode.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption related to telemetry attempts and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.This investigation will be updated should further pertinent information be provided.
 
Event Description
It was reported that this pacemaker was explanted as it transitioned into safety mode.The patient felt palpitations, light headedness and fell once due to dizzynes.A new non bsc device was implanted at the same surgical procedure.The pacemaker has been returned for analysis.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGENIO
Type of Device
IMPLANTABLE PACEMAKER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15137459
MDR Text Key296950704
Report Number2124215-2022-28337
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526536786
UDI-Public00802526536786
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N970003/S132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/01/2014
Device Model NumberK174
Device Catalogue NumberK174
Device Lot Number111472
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age51 YR
Patient SexFemale
-
-