Model Number K174 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problems
Fall (1848); Dizziness (2194); Presyncope (4410); Unspecified Heart Problem (4454)
|
Event Date 07/09/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
This investigation will be updated should further pertinent information be provided.
|
|
Event Description
|
It was reported that this pacemaker was explanted as it transitioned into safety mode.The patient felt palpitations, light headedness and fell once due to dizziness.A new non bsc device was implanted at the same surgical procedure.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode due to system resets.It was confirmed that brady therapy remained available.The system resets occurred during a telemetry session and caused the device to enter safety mode.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption related to telemetry attempts and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.This investigation will be updated should further pertinent information be provided.
|
|
Event Description
|
It was reported that this pacemaker was explanted as it transitioned into safety mode.The patient felt palpitations, light headedness and fell once due to dizzynes.A new non bsc device was implanted at the same surgical procedure.The pacemaker has been returned for analysis.No additional adverse patient effects were reported.
|
|
Search Alerts/Recalls
|