Model Number 16-02-80
Device Problem Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/04/2022
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in glasgow, united kingdom.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-cooler system 3t failed the electrical safety test due to high earth leakage current.There was no patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: through follow-up communication livanova learned that after a couple of weeks from the installation in june 2022 heater cooler 3t system started to generate alarms on their theatre monitoring system.At that point the unit failed the insulation test.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The failure was traced back to a defective heater element on the cardioplegia circuit.The part was replaced and the problem could be solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Complaint database review didn't identify short circuit events related to early failure of electrical safety test failure for high earth leakage current in last 3 years.Based on above, and considering that unit was manufactured in 2022, it cannot be ruled out that the reported issue was most likely due to heater element early failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key15138186
MDR Text Key298046382
Report Number9611109-2022-00374
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817901099
UDI-Public010403381790109911220519
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/01/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/19/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-