C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP M.R.I. IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 1806050J |
Device Problem
Difficult or Delayed Separation (4044)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/08/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a port placement procedure, air-guard was allegedly not divided well and the sheath was unable to be peeled away well.There was no reported patient injury.
|
|
Manufacturer Narrative
|
The catalog number identified has not been cleared in the us but is similar to the powerport isp m.R.I.Implantable port, chronoflex single-lumen, kit, 6f that are cleared in the us.The pro code and 510 k number for the powerport isp m.R.I.Implantable port, chronoflex single-lumen, kit, 6f are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 04/2023).Device not returned.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the powerport isp m.R.I.Implantable port, chronoflex single-lumen, kit, 6f that are cleared in the us.The pro code and 510 k number for the powerport isp m.R.I.Implantable port, chronoflex single-lumen, kit, 6f are identified in d2 and g4.H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 04/2023), g3, h6(method).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|
Event Description
|
It was reported that during a port placement procedure, air-guard was allegedly not divided well and the sheath was unable to be peeled away well.There was no reported patient injury.
|
|