Model Number G52918
Device Problem Difficult or Delayed Activation (2577)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/27/2022
Event Type  Injury  
Event Description
Description of event according to initial reporter: the physician placed filter, no deviation from how he always places them, from the right femoral vein approach.Message was that the red ¿lock¿ on the ivc filter wasn¿t working properly.The filter would not deploy and then it ended up deploying in the sheath.They tried to resheath it since it was only partially out of the sheath, however that just messed the secondary struts up.Ultimately, the physician got internal jugular access and had to snare the sheath to hold it in place while the tech slowly retracted the deployment sheath.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key15159101
MDR Text Key297211002
Report Number3005580113-2022-00087
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002529189
UDI-Public(01)10827002529189(17)250218(10)E4225464
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG52918
Device Catalogue NumberIGTCFS-65-1-UNI-TULIP
Device Lot NumberE4225464
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA08/03/2022
Distributor Facility Aware Date07/13/2022
Event Location Hospital
Date Report to Manufacturer08/03/2022
Date Device Manufactured02/18/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-