Catalog Number UNK HIP FEMORAL SLEEVE ASR |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Osteolysis (2377); Metal Related Pathology (4530)
|
Event Date 07/20/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that patient had original implant date unknown.Patient has bilateral asr on corail stem.On (b)(6) we revised the patients right asr due to cup being vertical.Radiographic signs of osteolysis.Cobalt serum levels were 80, chromium serum levels were 40.Showing signs of cobaltism.A competitor acetabular cup and dual mobility liner with a depuy 28mm ceramic +5 ts head was implanted.Doctor plans on revising patients left asr in october.No other information is available.Doi: unknown.Dor: (b)(6) 2022.Right hip.
|
|
Manufacturer Narrative
|
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: device history reviews for asr platform have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
|
|