• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 UNK HIP FEMORAL SLEEVE ASR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 UNK HIP FEMORAL SLEEVE ASR Back to Search Results
Catalog Number UNK HIP FEMORAL SLEEVE ASR
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Osteolysis (2377); Metal Related Pathology (4530)
Event Date 07/20/2022
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that patient had original implant date unknown.Patient has bilateral asr on corail stem.On (b)(6) we revised the patients right asr due to cup being vertical.Radiographic signs of osteolysis.Cobalt serum levels were 80, chromium serum levels were 40.Showing signs of cobaltism.A competitor acetabular cup and dual mobility liner with a depuy 28mm ceramic +5 ts head was implanted.Doctor plans on revising patients left asr in october.No other information is available.Doi: unknown.Dor: (b)(6) 2022.Right hip.
 
Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: device history reviews for asr platform have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK HIP FEMORAL SLEEVE ASR
Type of Device
FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15164054
MDR Text Key297234816
Report Number1818910-2022-15063
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL SLEEVE ASR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
ASR UNI FEMORAL IMPL SIZE 46; UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL STEM CORAIL
Patient Outcome(s) Required Intervention;
Patient Age81 YR
Patient SexFemale
-
-